SecurityContext

CVE-2015-3953: Hospira Infusion Systems - Hardcoded Credentials

  • Cve: CVE-2015-3953
  • Name: Hospira Infusion Systems - Hardcoded Credentials
  • Description: Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
  • Impact: Attackers can gain unauthorized access, potentially leading to data tampering, system control, or disruption of medical device operation.
  • Remediation: Close Port 20/FTP and Port 23/TELNET on affected devices; update to Plum 360 Infusion System which is not vulnerable.
  • Vulnerability type: hardcoded_credentials
  • Vulnerability impact: privilege_escalation, information_disclosure, data_tampering
  • Requirement type: none
  • Labels source: projectdiscovery
  • Severity: critical
  • Cvss score: 9.8
  • Cvss vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
  • Epss score: 0.01972
  • Epss percentile: 0.80987
  • Assignee: [email protected]
  • Published at: 2019-03-25T17:29:00Z
  • Age in days: 2801
  • Modified at: 2026-06-17T00:26:32Z
  • References total: 1
  • Products total: 3
  • Advisory url: https://nvd.nist.gov/vuln/detail/CVE-2015-3953
  • Nvd status: modified
  • Checked, none or zero: patch available, known exploited entries, nuclei templates, public exploits total, public exploits capped, public exploits, oss, projects total, projects, rejected by nvd
  • Not known: requirements, hackerone

Cwe

IdName
CWE-259Use of Hard-coded Password
CWE-798Use of Hard-coded Credentials

References

UrlSourceTags
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01nvdMitigation, Third Party Advisory, US Government Resource

Products

Same in every row:

  • Platform bound: yes
  • Sources: nvd
ProductAffectedUrl
pifzer:plum_a+3_infusion_system_firmware<= 13.6/v1/products/pifzer/plum_a+3_infusion_system_firmware
pifzer:plum_a+_infusion_system_firmware<= 13.4/v1/products/pifzer/plum_a+_infusion_system_firmware
pifzer:symbiq_infusion_system_firmware<= 3.13/v1/products/pifzer/symbiq_infusion_system_firmware

Index

  • Checked, none or zero: repositories, repositories high signal, packages

Exploitation

  • Remote: yes
  • Checked, none or zero: known exploited, known exploited by, exploitation seen, public exploit, public exploit count, initial access, authentication required
  • Not known: known exploited since, vulncheck known exploited since, public exploit first date, nuclei template, initial access since, hackerone reports, exposed hosts