CVE-2019-6538: Medtronic Conexus Protocol - Broken Access Control
- Cve: CVE-2019-6538
- Name: Medtronic Conexus Protocol - Broken Access Control
- Description: The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
- Impact: Attackers can manipulate device memory, potentially causing device malfunction or harm to patients.
- Remediation: Implement proper authentication and authorization mechanisms in the telemetry protocol; update device firmware if available.
- Vulnerability type: misconfiguration
- Vulnerability impact: information_disclosure, data_tampering
- Requirement type: user_interaction
- Requirements: Attacker needs short-range access when device radio is on
- Labels source: projectdiscovery
- Severity: medium
- Cvss score: 6.5
- Cvss vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
- Epss score: 0.00895
- Epss percentile: 0.60453
- Assignee: [email protected]
- Published at: 2019-03-25T22:29:00Z
- Age in days: 2801
- Modified at: 2026-06-17T02:39:12Z
- References total: 2
- Products total: 20
- Advisory url: https://nvd.nist.gov/vuln/detail/CVE-2019-6538
- Nvd status: modified
- Checked, none or zero: patch available, known exploited entries, nuclei templates, public exploits total, public exploits capped, public exploits, oss, projects total, projects, rejected by nvd
- Not known: hackerone
Cwe
| Id | Name |
|---|---|
| CWE-284 | Improper Access Control |
| CWE-306 | Missing Authentication for Critical Functionality |
References
| Url | Source | Tags |
|---|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01 | nvd | Mitigation, Third Party Advisory, US Government Resource |
| http://www.securityfocus.com/bid/107544 | nvd | Third Party Advisory, VDB Entry |
Products
Same in every row:
- Affected: all versions
- Platform bound: yes
- Sources: nvd
Index
- Checked, none or zero: repositories, repositories high signal, packages
Exploitation
- Checked, none or zero: known exploited, known exploited by, exploitation seen, public exploit, public exploit count, initial access, remote, authentication required
- Not known: known exploited since, vulncheck known exploited since, public exploit first date, nuclei template, initial access since, hackerone reports, exposed hosts