  • Cve: CVE-2019-6538
  • Name: Medtronic Conexus Protocol - Broken Access Control
  • Description: The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
  • Impact: Attackers can manipulate device memory, potentially causing device malfunction or harm to patients.
  • Remediation: Implement proper authentication and authorization mechanisms in the telemetry protocol; update device firmware if available.
  • Vulnerability type: misconfiguration
  • Vulnerability impact: information_disclosure, data_tampering
  • Requirement type: user_interaction
  • Requirements: Attacker needs short-range access when device radio is on
  • Labels source: projectdiscovery
  • Severity: medium
  • Cvss score: 6.5
  • Cvss vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
  • Epss score: 0.00895
  • Epss percentile: 0.60453
  • Assignee: [email protected]
  • Published at: 2019-03-25T22:29:00Z
  • Age in days: 2801
  • Modified at: 2026-06-17T02:39:12Z
  • References total: 2
  • Products total: 20
  • Advisory url: https://nvd.nist.gov/vuln/detail/CVE-2019-6538
  • Nvd status: modified
  • Checked, none or zero: patch available, known exploited entries, nuclei templates, public exploits total, public exploits capped, public exploits, oss, projects total, projects, rejected by nvd
  • Not known: hackerone

Cwe

IdName
CWE-284Improper Access Control
CWE-306Missing Authentication for Critical Functionality

References

UrlSourceTags
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01nvdMitigation, Third Party Advisory, US Government Resource
http://www.securityfocus.com/bid/107544nvdThird Party Advisory, VDB Entry

Products

Same in every row:

  • Affected: all versions
  • Platform bound: yes
  • Sources: nvd
ProductUrl
medtronic:amplia_crt-d_firmware/v1/products/medtronic/amplia_crt-d_firmware
medtronic:carelink_2090_firmware/v1/products/medtronic/carelink_2090_firmware
medtronic:claria_crt-d_firmware/v1/products/medtronic/claria_crt-d_firmware
medtronic:compia_crt-d_firmware/v1/products/medtronic/compia_crt-d_firmware
medtronic:concerto_crt-d_firmware/v1/products/medtronic/concerto_crt-d_firmware
medtronic:concerto_ii_crt-d_firmware/v1/products/medtronic/concerto_ii_crt-d_firmware
medtronic:consulta_crt-d_firmware/v1/products/medtronic/consulta_crt-d_firmware
medtronic:evera_icd_firmware/v1/products/medtronic/evera_icd_firmware
medtronic:mirro_icd_firmware/v1/products/medtronic/mirro_icd_firmware
medtronic:nayamed_nd_icd_firmware/v1/products/medtronic/nayamed_nd_icd_firmware