SecurityContext

CVE-2020-27252: Medtronic MyCareLink Smart - Race Condition

  • Cve: CVE-2020-27252
  • Name: Medtronic MyCareLink Smart - Race Condition
  • Description: Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
  • Impact: Attackers can remotely execute arbitrary code on the device, potentially leading to full control and manipulation of the device's functions.
  • Remediation: Implement proper firmware signature verification and update to the latest software version.
  • Vulnerability type: race_condition
  • Vulnerability impact: remote_code_execution
  • Requirement type: none
  • Labels source: projectdiscovery
  • Severity: high
  • Cvss score: 8.1
  • Cvss vector: CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
  • Epss score: 0.0384
  • Epss percentile: 0.90387
  • Assignee: [email protected]
  • Published at: 2020-12-14T20:15:12Z
  • Age in days: 2171
  • Modified at: 2026-06-17T03:09:03Z
  • References total: 3
  • Products total: 1
  • Advisory url: https://nvd.nist.gov/vuln/detail/CVE-2020-27252
  • Nvd status: modified
  • Checked, none or zero: patch available, known exploited entries, nuclei templates, public exploits total, public exploits capped, public exploits, oss, projects total, projects, rejected by nvd
  • Not known: requirements, hackerone

Cwe

IdName
CWE-367Time-of-check Time-of-use (TOCTOU) Race Condition

References

Same in every row:

  • Source: nvd
UrlTags
https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01Third Party Advisory, US Government Resource
https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-smart-security-vulnerability-patch.html
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-345-01

Products

ProductAffectedPlatform boundSourcesUrl
medtronic:mycarelink_smart_model_25000_firmwareall versionsyesnvd/v1/products/medtronic/mycarelink_smart_model_25000_firmware

Index

  • Checked, none or zero: repositories, repositories high signal, packages

Exploitation

  • Remote: yes
  • Checked, none or zero: known exploited, known exploited by, exploitation seen, public exploit, public exploit count, initial access, authentication required
  • Not known: known exploited since, vulncheck known exploited since, public exploit first date, nuclei template, initial access since, hackerone reports, exposed hosts